Blank standard does not show up in the Empower calibration curve - WKB119942
Article number: 119942
SYMPTOMS
- Standard concentrations of 0, 25, 50, 75, and 100
- The 0 standard does not show up in the curve
- Objective is to have the calibration curve go through zero
ENVIRONMENT
- Empower
CAUSE
0 standard does not have a peak, so it will be missing when integrated. Therefore, it does not show up in the calibration curve.
FIX or WORKAROUND
The best way to have the calibration curve go through 0 is to use Linear thru Zero fit in the processing method.