Incorrect amounts are calculated for unknowns - WKB119227

Article number: 119227

SYMPTOMS

  • Incorrect amounts are calculated for unknowns

ENVIRONMENT

  • Empower 3

CAUSE

The incorrect standard curve was associated with the processing method being used to process the unknown samples.

FIX or WORKAROUND

  1. Process the correct standards first, making sure to clear the existing calibration.
  2. Process the unknowns.

ADDITIONAL INFORMATION

 