SYMPTOMS
- Error in the Empower message center: "No results produced for this processing job"
- No results produced after batch processing using specified processing method
- More than 100 MB free project tablespace
ENVIRONMENT
- Empower 3
- 2996 PDA detector
- ACQUITY TQD
CAUSE
3D data needs a method set for processing and cannot process using a processing method.
FIX or WORKAROUND
- Process the sample set using a specified method set.
ADDITIONAL INFORMATION
id11730, 2996, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP