No results produced with 3D data using Empower - WKB11730

Article number: 11730

SYMPTOMS

  • Error in the Empower message center: "No results produced for this processing job"
  • No results produced after batch processing using specified processing method
  • More than 100 MB free project tablespace

ENVIRONMENT

  • Empower 3
  • 2996 PDA detector 
  • ACQUITY TQD

CAUSE

3D data needs a method set for processing and cannot process using a processing method.

FIX or WORKAROUND

  1. Process the sample set using a specified method set. 

ADDITIONAL INFORMATION