Equilibrate and condition column using Agilent LC cannot run - WKB115822

Article number: 115822

SYMPTOMS

  • Equilibrate or condition column using Agilent LC in sample set last row cannot run

ENVIRONMENT

  • Empower 3 
  • Agilent ICF Support layer
  • Agilent Infinity 1260 LC

CAUSE

 The equilibrate or condition column instrument method is incorrect, because the the Shutdown after Run checkbox is selected

 

FIX or WORKAROUND

Uncheck the Shutdown after Run checkbox on the instrument method General table and save the instrument method.

ADDITIONAL INFORMATION

Do not use the LC Shutdown Method in Sample set for regular analysis. Instead, specify this method as the Shutdown method on the Run Samples Defaults screen found under the Customize/Defaults menu selection of Run Samples or QuickStart.

 