With Empower 3 and a 2998 PDA, how to get data by subtracting the reference wavelength from a specific wavelength - WKB115706

Article number: 115706

OBJECTIVE or GOAL

  • Acquire data obtained by subtracting the reference wavelength from a specific wavelength.

ENVIRONMENT

  • Empower 3
  • 2998 Photodiode Array Detector

PROCEDURE

  1. Open the "2D Channel" tab in the instrument method edit screen of the 2998 detector.
  2. In the Channel 1 check box, select "Difference" from the "Data Mode" drop-down list.
  3. Specify the two wavelengths.
    clipboard_edf129d8bf780ba0246408661a6690fa1.png
  4. Save the device method and perform the measurement.

 

ADDITIONAL INFORMATION