Skip to main content
Waters

Agilent ELSD G7102A does not go to next injection - WKB114552

Article number: 114552

SYMPTOMS

  • After the first injection in a sample set, the Agilent LC does not go to next injection
  • System configured properly through preconfiguration
  • System communicates normally
  • Collects Agilent ELSD signal only
  • Agilent ELSD meets the minimum firmware requirement

ENVIRONMENT

  • Empower 3
  • ICF 3.1
  • Agilent 1290 Infinity 2
  • Agilent ELSD G7102A

CAUSE

The instrument method is set up to collect only the ELSD channel.

FIX or WORKAROUND

Add an additional channel like the PDA or one of the AUX channels like pressure. This will allow the system to move to the next injection because it is not waiting for a trigger.

ADDITIONAL INFORMATION

 

id114552, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPELS, UPELSARC

Not able to find a solution? Click here to request help.