"Error enabling StartTLS encryption with server" when attempting to configure LDAP authentication in Empower 3 FR5 - WKB110163

Article number: 110163

SYMPTOMS

  • The following error messages appear when logging in to or testing the LDAP connection parameters in Empower:
    • Error enabling StartTLS encryption with server 'ldap-server'
    • Connection to LDAP server 'ldap-server' on port XXXX failed.
    • Server is not willing to handle directory requests.

ENVIRONMENT

  • Empower 3 Feature Release 5
  • LDAP authentication

CAUSE

An alias is used to route LDAP authentication requests to a load balancer. The load balance does not support StartTLS.

FIX or WORKAROUND

  1. Discontinue use of the alias and / or load balancer.
  2. Configure Empower to point to specific directory servers to service LDAP authentication requests.

ADDITIONAL INFORMATION

 