Run does not proceed to the next injection with Agilent ALS, Pump and e-SAT/IN - WKB10864
Article number: 10864
SYMPTOMS
- First injection begins, however, it does not end (does not proceed to the next injection)
- At the end of the run, system continues to show that it is acquiring for 10 minutes then an Instrument failure message appears
ENVIRONMENT
- Empower 3 Enterprise SR2 HF1
- Agilent ICF Support v2.2 with ICF A.02.03 DU2
- Chromatographic System:
- Agilent 1100 ALS (controlled by Empower)
- Agilent 1100 Pump (controlled by Empower)
- e-SAT/IN (controlled by Empower)
- Dionex ICS 2000 integrated IC (controlled by Chromeleon)
- Remote start cable connected from Agilent pump to Dionex IC
- Inject start cable connected from Dionex IC to e-SAT/IN
CAUSE
- Suspect intermittent start signal to Dionex IC (and thereby, to the e-SAT/IN)
FIX or WORKAROUND
- Reconfigure the system so that the start/stop signal is coming from the Agilent autosampler to the e-SAT/IN.
- Enable an auxiliary channel on the Agilent pump in the instrument method (auxilliary channel will ensure that the run ends).