Cannot log into Empower due to ORA-12514 Oracle error - WKB107673
Article number: 107673
SYMPTOMS
- Cannot log into Empower due to ORA-12514 Oracle error
ENVIRONMENT
- Empower
CAUSE
Failed Raw Data hard drive. When Oracle starts up, it looks for previously configured folders for mirror and archivedb folders on this drive. If they are missing, then the error is displayed and you can't log into Empower
FIX or WORKAROUND
- Check Windows Event Viewer system log for disk errors:
- Replace failed hard drive
- Restore folders and contents to it
- Restart the server or database
ADDITIONAL INFORMATION
Maintain uniformity while restoring the folders and its structure