Instrument method does not appear in the method drop-down list in the Run Samples window in Empower - WKB102223
Article number: 102223
SYMPTOMS
- Instrument Method does not appear in the method drop-down list in the Run Samples window
- The instrument method is on the project's Methods tab
ENVIRONMENT
- Empower 3
CAUSE
The instruments in the instrument method do not match the instruments that are part of the chromatographic system that is currently being used by the open Run Samples window.
FIX or WORKAROUND
Adjust the instrument method's list of instruments to match the system configuration being used by the Run Samples window.