Empower 3 does not see the 2707 Autosampler - WKB10049

Article number: 10049

SYMPTOMS

  • 2707 does not show in the node properties
  • DHCP shows the IP address but no model
  • Unable to see the 2707 using the firmware loader
  • 2707 responds to a ping

ENVIRONMENT

  • 2707 Autosampler
  • Empower 3 FR3
  • 2707 ICS v1.20
  • 2707 Firmware 1.15

CAUSE

2707 Autosampler firmware was higher than the one that comes with the ICS loaded

FIX or WORKAROUND

  1. Upgrade 2707 ICS to the current version v1.30