How can I process four liters (4L) of water through SPE cartridges? - WKB73221

Article number: 73221

ENVIRONMENT

  • Solid-phase extraction (SPE)
  • High-volume samples
  • Vacuum manifold
  • Oasis
  • Sep-Pak

ANSWER

For 1-, 3-, and 6-cc cartridges, use:
WAT054260 - SEP-PAK RESERVOIR ADAPTOR
WAT024310 - MALE/MALE LUER FITTING 100/BX

For 5-cc glass cartridges, use:
405000934 - ADAPTOR, 5cc, Teflon 10/pkg (Waters does not have a Teflon male-male adaptor.)

For 12-, 20-, and 35-cc cartridges, use:
WAT048160 - ADAPTER,12,20,35CC
WAT024310 - MALE/MALE LUER FITTING 100/BX

Plug a male-male adapter into the adapter opening.

Connect 1/8 inch inner diameter (ID) tubing to the male-male adapter and run the tubing into the sample container.

Examples:

clipboard_eb2373123a4cc3910696e7df2cb9ea870.png

ADDITIONAL INFORMATION

See the Sample Preparation chapter in the current catalog to learn more.