What SPE product could offer a simple means of extracting nucleotides, nucleosides, and nucleobases from biological samples? - WKB69991
Article number: 69991
ENVIRONMENT
- Solid-phase extraction (SPE)
- Nucleotides
- Nucleosides
- Nucleobases
- Oasis HLB
- Sep-Pak Accell Plus QMA
ANSWER
Oasis HLB would be a good starting point.
Load the samples in 100% aqueous, and then elute in methanol. An effective wash step could also be added, being careful not to elute any of the nucleotides/nucleosides/nucleobases of interest.
Use the Waters SPE optimization procedures listed in the care and use manual for improving recovery.
This procedure simultaneously helps with desalting.
ADDITIONAL INFORMATION
For additional information, see the Oasis HLB Cartridges and 96-Well Plates Care & Use Manual.
Sep-Pak QMA is an alternative. The compounds could be extracted using ion exchange.
id69991,