Can Ostro plates be used as flow-through SPE plates to capture phospholipids from a solution containing proteins, so that the phospholipids and the proteins can be analyzed separately? - WKB62521
Article number: 62521
ENVIRONMENT
- Ostro phospholipid removal plates
- Protein precipitation plate
- Sample preparation
- Solid-phase extraction
ANSWER
No.
ADDITIONAL INFORMATION
At least 50% acetonitrile is required for phospholipids to retain on the Ostro plate.
Proteins will precipitate under these conditions.
See the following related application note: Extraction of Phospholipids from Plasma Using Ostro Sample Preparation