What Oasis product is a good choice for extraction of porphyrins from tissue? - WKB60143
Article number: 60143
ENVIRONMENT
- Sample preparation
- Solid-phase extraction (SPE)
- Oasis cartridges
ANSWER
Although Waters has not evaluated Oasis products for the extraction of porphyrins, the following may be suitable in this application:
- Oasis PRiME HLB
- Oasis HLB
- Oasis PRiME MCX
- Oasis MCX
ADDITIONAL INFORMATION
When loading tissue extracts, ensure that the organic concentration is low (<10%) for loading, because too much organic in the loading sample solvent can cause analyte to elute during. Dilution of the extract with aqueous solution may be performed prior to sample loading.
Oasis Cartridges and 96-Well Plates Care & Use Manual
id60143,