Does Waters have a method for quantitating polysorbate? - WKB229757

Article number: 229757

ENVIRONMENT

  • Quantitation
  • Polysorbate

ADDITIONAL INFORMATION

The methods above use the Oasis MAX Online Column, 80Å, 30 µm, 2.1 mm X 20 mm (186002052).

  • Polysorbate is trapped at low pH while the therapeutic and other excipients are unretained.
    • Polysorbate 80 is a nonionic surfactant.
    • Polysorbate 80 is retained onto Oasis MAX by reversed phase only.
  • Unretained components are diverted to waste via a switching valve.
    • At low pH, biomolecules will carry a positive charge.
      • The positive charge on the Oasis MAX particles will repel the positively charged biomolecule off of the packing.
    • Large biomolecules are excluded from the pores of the Oasis MAX packing.
  • Polysorbate is eluted from the column to the ELSD detector when the switching valve switches back for elution.

The proposed workflow represents a straightforward and efficient method for analyzing polysorbate content in biotherapeutic formulations.

In some cases, the actual gradient table may need to be adjusted in order to properly allow the elution of biomolecules before the polysorbate is able to be eluted away from the biomolecules.

Related:

Determination of Fatty Acids Composition in Polysorbate 20 Pharmaceutical Raw Material Using HPLC with MS Detection

Determination of Fatty Acids in Polysorbate 80 Pharmaceutical Raw Materials by HPLC With Mass Detection

Quantitative Analysis of Polysorbate 20/80 in Protein-Based Biopharmaceuticals Using A One-Pot RPLC-MS Based Platform Method

An Efficient UV-Based Method for the Assessment of Oleic Acid Content in Biotherapeutic Drug Products Containing Polysorbate-80

 

id229757,

Not able to find a solution? Click here to request help.