What is an SPE recommendation for cleanup to analyze glycerol from fermentation sample? - WKB224208

Article number: 224208

ENVIRONMENT

  • Solid-phase extraction (SPE)
  • Method development
  • Glycerol

ANSWER

Sep-Pak C18, Sep-Pak Silica, and Sep-Pak NH2 are sorbent options to try.

ADDITIONAL INFORMATION

The Sep-Pak C18 would not retain the glycerol, and it could be used as a possible pass-through option to collect the effluent and retain the hydrophobic interferences.

The Silica and NH2 would retain the glycerol in a normal phase and/or HILIC manner and could be used to develop a method to remove sugars from the glycerol.