Why might I be getting poorly reproducible recovery for serine when doing a microwave hydrolysis? - WKB203619
Article number: 203619
ENVIRONMENT
- AccQ•Tag
- Serine
- Threonine
- Microwave hydrolysis
ANSWER
Serine and threonine are sensitive amino acids that degrade in standard hydrolysis conditions.
ADDITIONAL INFORMATION
Time-dependent studies are typically done to optimize hydrolysis conditions for serine and threonine, independent of the standard hydrolysis conditions.
Optimal microwave hydrolysis conditions must be investigated for each individual protein/peptide sample.
For additional information, see USP General Chapter 1052 Biotechnology-derivized Amino Acid Analysis.
And journal articles:
- Fred Downs and Ward Pigman, "The Destruction of Serine and Threonine during Acid Hydolysis", International Journal of Protein Research I, (1969), 181-184
- Michael Fountoulakis and Hans-Werner Lahm, "Review: Hydrolysis and amino acid composition analysis of proteins, Journal of Chromstography A, 826 (1998) 109-134