Does Waters sell an adapter to connect the blue tubing from the Positive Pressure-96 Processor (186006961) to a house nitrogen source spigot (barb)? - WKB109317

Article number: 109317

ENVIRONMENT

  • Positive Pressure-96 Processor
  • Solid-phase extraction (SPE)

ANSWER

No

ADDITIONAL INFORMATION

The tubing that comes with the Waters Positive Pressure-96 Processor (186006961) is not designed to push directly onto a house nitrogen source spigot (barb).  The tubing may pop off of the spigot (barb) during use.

Instead, connect to a pressure regulator and then connect the pressure regulator to the nitrogen source using appropriate tubing and fittings to make sure the tubing is securely fit onto the nitrogen source spigot (barb). 

See the Waters Positive Pressure-96 Processor Operator’s Guide for more information.