What vials are recommended for PFC analysis solution? - WKB79128
Article number: 79128
ENVIRONMENT
- vial
- PFC
- PFOA
- PFOS
- PFAS
- septumless
ANSWER
SKU: 186004112 - Polypropylene 12 x 32 mm Screw Neck Vial, with Polyethylene Septumless Cap, 300 µL Volume, 100/pk
or
SKU: 186005230 - Polypropylene 12 x 32 mm Screw Neck Vial, with Polyethylene Septumless Cap, 700 µL Volume, 100/pk
ADDITIONAL INFORMATION
In some cases, if Polyethylene Septumless caps are not preferred:
A Waters scientist has used SKU: 186000305 - Blue, 12 x 32 mm Screw Neck Cap and Preslit PTFE/Silicone Septum, 100/pk and had not seen seen PFAS contamination when using them. The scientist had tested PFAS down to sub-ppt levels when using this cap and septa.
Each Laboratory would need to verify if this cap/septa assemply leads to any amount of PFAS contamination through experimental studies.
id79128,