How do I select an Acrodisc filter? - WKB77649
ENVIRONMENT
- Acrodisc filters
ANSWER
Select filters on the basis of four main criteria:
1. The nature of the sample:
Aqueous: GHP, nylon, PVDF, PES
Nonpolar: CR/GHP, PTFE, GHP
Protein: GHP, PVDF
Ion chromatography: IC, PES
2. What size are the particles in the column?
Greater than 3 µm: Use 0.45-µm filter pore.
Less than 3 µm: Use 0.2-µm filter pore.
3. What volume is your sample?
Less than 2 mL: Use 4-mm Acrodisc. Hold up volume less than 10 µL.
Less than 10 mL: Use 13-mm minispike Acrodisc. Hold up volume less than 14 µL.
Less than 10 mL: Use 13-mm male luer Acrodisc. Hold up volume less than 30 µL.
Less than 100 mL: Use 25-mm Acrodisc. Hold up volume less than 100 µL.
4. Use the solvent compatibility chart to determine if the solvent used is compatible with the solvent being run through the cartridge.
ADDITIONAL INFORMATION
Online Filter Selector Tool: http://www.waters.com/app/selector/en/filters.html