Does the PFC (perfluorinated compounds) analysis kit part number 205000588 contain chemistry consumable items? - WKB66434

Article number: 66434

ENVIRONMENT

  • PFC
  • PFOA
  • PFOS

ANSWER

No. 

The Perfluorinated compounds analysis kit, part number 205000588, contains only the hardware components necessary to minimize contamination from the UPLC and mobile phase that would interfere with PFAS (perfluoroalkyl and polyfluoroalkyl substances) analysis.

The kit contains the 50uL extension loop, stainless steel tubing to accommodate the isolator column, and the PEEK solvent lines to replace the standard solvent inlet lines.  The chemistries necessary will vary based on which application is being run. 

For application specific chemistry ordering refer to the PFAS analysis ordering information guide https://www.waters.com/webassets/cms...20007102en.pdf.   

ADDITIONAL INFORMATION

For additional information visit https://www.waters.com/waters/en_US/...6&locale=en_US.

Not able to find a solution? Click here to request help.