Skip to main content
Waters

Can the ARC Premier be exposed to sodium hydroxide? - WKB232511

Article number: 232511

ENVIRONMENT

  • ARC Premier
  • Sodium hydroxide (NaOH)

ANSWER

The ARC Premier should not be exposed to sodium hydroxide above pH 10.

ADDITIONAL INFORMATION

The MaxPeak High Performance Surfaces will degrade above pH 10. Stay within the system pH limit.

For additional information, see the ARC Premier System Guide.