What is the recommended storage solvent for the XBridge BEH Amide and ACQUITY BEH Amide columns? - WKB99387
Article number: 99387
ENVIRONMENT
- Column care
- XBridge BEH Amide
- ACQUITY BEH Amide
- storage solvent
ANSWER
Use 100% ACN when storing the XBridge BEH Amide or the ACQUITY BEH Amide for longer than four (4) days.
ADDITIONAL INFORMATION
If in continual use and not sitting unused for more than four days, the XBridge BEH Amide and ACQUITY BEH Amide may be stored in 90:10 acetonitrile/water.
The use of 10% water at the end of the day in a shutdown method allows for faster equilibration of the column in the initial mobile phase for the next day's use.
For additional information see:
XBridge Columns Care & Use Manual (720001366EN)
ACQUITY UPLC BEH Columns Care & Use Manual (715001371) (includes BEH Amide & BEH HILIC)
id99387, eluent, UPBINARY