Which column should I use for thiamine diphosphate analysis? - WKB98172

Article number: 98172

ENVIRONMENT

  • Atlantis
  • PREMIER
  • Bridged Ethylene Hybrid (BEH)
  • C18
  • Anion Exchange (AX)
  • Mixed-mode
  • Polar retention
  • Phosphorylated compounds

ANSWER

Atlantis PREMIER BEH C18 AX columns.

ADDITIONAL INFORMATION

These columns can retain by both reversed-phase and anion exchange retention mechanisms.  

Run in a mobile phase buffered to a pH at least 2 pH units above the weakest acidic pKa.

 Atlantis PREMIER Columns Infographic

Atlantis Premier BEH C18 AX Columns Care and Use Manual

A New Mixed-Mode Reversed-Phase/Anion-Exchange Stationary Phase Based on Hybrid Particles