Can I run Atlantis PREMIER BEH C18 AX VanGuard FIT Column, 2.5 µm, 4.6 x 150 mm, 1/pk (186009385) on a traditional HPLC with a maximum pressure limit of 4,000 psi? - WKB97656
ENVIRONMENT
- High Performance Liquid Chromatography (HPLC)
- Ultra High Performance Liquid Chromatography (UHPLC)
- Column
- Bridged Ethylene Hybrid (BEH)
- eXtended Performance (XP) 2.5 um
- Polar compound retention
- Reversed-Phase
- Mixed-Mode
- Anion Exchange
- PREMIER
- High Performance Surfaces (HPS)
- MaxPeak HPS
- Fully Integrated Technology (FIT)
- VanGuard
ANSWER
Yes.
ADDITIONAL INFORMATION
To use on a traditional HPLC system, the organic modifier should be acetonitrile and not methanol.
Use of methanol causes a significant increase in column back pressure as the methanol and water mix.
Use of acetonitrile helps keep pressures lower.
Heating the column also reduces back pressure.
You can estimate back pressures by using the Columns Calculator Version 2.0. (You may need approval from your IT department to download this calculator since it is an executable file).