What connector should I used to connect XBridge BEH SEC 7.8 mm ID columns in series? - WKB97346
Article number: 97346
ENVIRONMENT
- XBridge
- Column
- Bridged Ethylene Hybrid (BEH)
- Size exclusion Chromatography (SEC)
- Connectors
ANSWER
WAT084080 -Tubing Assembly, Column Joining (0.009" ID) - for lower pressure systems such as HPLC.
or
700009535 - ASSY, TUBE, .004 ID, COL. CONN., "U" - For higher pressure systems such as UHPLC and UPLC.
ADDITIONAL INFORMATION
700009535 caused higher back pressures, so this is not the first choice of connector when running on an HPLC system.