What are the cleaning recommendations for the Sugar-Pak 1 column (WAT085188)? - WKB92565

Article number: 92565

ENVIRONMENT

  • Sugar-Pak 1
  • Column cleaning

ANSWER

Reverse column flow once per day or every 500-1000 mL with an aqueous solution of 500 mg/L calcium disodium EDTA. To regenerate, use the same solution overnight at 90 °C and 0.5 mL/min in the reverse position. Reverse the column back to normal orientation when you resume operation.

ADDITIONAL INFORMATION

For additional information, see the Sugar-Pak I Column Care and Use Manual.