Can I use nanoEase M/Z columns with ZenFit fittings on a nanoACQUITY system? - WKB87468
Article number: 87468
ENVIRONMENT
- nanoEase M/Z
- ZenFit
- nanoACQUITY
ANSWER
Yes
ADDITIONAL INFORMATION
ZenFit fittings will fit into the V-detail (10-32 thread, 1/16-inch OD tubing) fittings on a nanoACQUITY system.
For trapping:
ZenFit fittings will not fit into the M-detail nanoACQUITY nanoTee. (M-detail means a 6-40 thread and a 1/32-inch OD tubing.)
When using nanoEase traps and columns with ZenFit fittings, you can use the following:
700011518 - ZenFit Assy, Tee
This will accept ZenFit fittings only.
Waters does not have a ZenFit Tee holder for nanoACQUITY systems, so this ZenFit Tee would simply sit inside the column heater, unattached.
nanoEase M/Z Columns Ordering Information
ZenFit Connection Technology and nanoEase M/Z columns Brochure
nanoEase M/Z Columns Care and Use Manual