What is the loading capacity of BEH SEC columns? - WKB76144
Article number: 76144
ENVIRONMENT
- Size Exclusion Chromatography (SEC)
- Bridged Ethylene Hybrid (BEH)
- ACQUITY
- XBridge
ANSWER
Mass Load:
- <100 µg for a 4.6x150mm ACQUITY UPLC BEH SEC Column.
- <200 µg for a 4.6x300mm ACQUITY UPLC BEH SEC Column.
- <100 µg for a 4.6x150mm XBridge BEH SEC Column.
- <200 µg for a 4.6x300mm XBridge BEH SEC Column.
- <300 µg for a 7.8x150mm XBridge BEH SEC Column.
- <600 µg for a 7.8x300mm XBridge BEH SEC Column.
Volume Load:
- <20 µL for a 4.6x150mm ACQUITY UPLC BEH SEC Column.
- <40 µL for a 4.6x300mm ACQUITY UPLC BEH SEC Column.
- <21 µL for a 4.6x150mm XBridge BEH SEC Column.
- <42 µL for a 4.6x300mm XBridge BEH SEC Column.
- <60 µL for a 7.8x150mm XBridge BEH SEC Column.
- <120 µL for a 7.8x300mm XBridge BEH SEC Column.
ADDITIONAL INFORMATION
Related links:
BEH SEC Column Technology – Infographic
ACQUITY UPLC BEH SEC Columns and Standards Care & Use Manual
XBridge Protein BEH SEC Columns & Standards Care & Use Manual
Newer SEC column technology (250A):