Are guards available for Protein-Pak Hi Res columns? - WKB73607
Article number: 73607
ENVIRONMENT
- Protein-Pak Hi Res columns
ANSWER
No, guards are not available for these columns.
ADDITIONAL INFORMATION
Use SKU 205000343 (ACQUITY UPLC Column In-Line Filter Kit) to protect the column.
NOTE:
The internal diameter of this filter unit is only 2.1 mm, which will contribute to higher pressure before the column.
The pore size in this filter is only 0.2 µm. Use caution to prevent microbial growth in the aqueous mobile phase, as microbes can plug this filter.