Skip to main content
Waters

What is a good column choice for the reversed-phase analysis of peptides? - WKB73119

Article number: 73119

ENVIRONMENT

  • UltraPerformance Liquid Chromatography (UPLC)
  • Ultra High Performance Liquid Chromatography (UHPLC)
  • High Performance Liquid Chromatography (HPLC)
  • Peptide separation technology (PST)
  • Peptide mapping
  • Peptide analysis
  • Proteomics
  • Column
  • ACQUITY
  • MaxPeak
  • High Performance Surfaces (HPS)
  • ACQUITY Premier
  • Bridged Ethylene Hybrid (BEH)
  • High Strength Silica (HSS)
  • Charged Surface Hybrid (CSH)
  • Sub-2um particle size
  • XBridge
  • XSelect

ANSWER

For UPLC

  • ACQUITY Premier Peptide BEH C18, 130Å, 1.7 µm Columns**
  • ACQUITY Premier Peptide BEH C18, 300Å, 1.7 µm Columns**
  • ACQUITY Premier Peptide CSH C18 Column, 130Å, 1.7 µm Columns**
  • ACQUITY Premier Peptide HSS T3 Column, 100Å, 1.8 µm Columns**
  • ACQUITY UPLC Peptide BEH C18, 130Å, 1.7µm columns
  • ACQUITY UPLC Peptide BEH C18, 300Å, 1.7µm columns
  • ACQUITY UPLC Peptide CSH C18, 130Å, 1.7µm columns
  • ACQUITY UPLC Peptide HSS T3, 100Å, 1.8µm columns

**MaxPeak HPS (High Performance Surfaces) column technology.

For UHPLC and HPLC:

  • XBridge Premier Peptide BEH C18, 130Å, 2.5 µm Columns**
  • XBridge Premier Peptide BEH C18, 300Å, 2.5 µm Columns**
  • XSelect Premier Peptide CSH C18, 130Å, 2.5 µm Columns**
  • XSelect Premier Peptide HSS T3, 100Å, 2.5 µm Columns**
  • XBridge Peptide BEH C18, 130Å, 2.5µm XP, 3.5µm, 5µm, and 10µm Columns
  • XBridge Peptide BEH C18, 300Å,  2.5µm XP, 3.5µm, 5µm, and 10µm Columns
  • XSelect Peptide CSH C18, 130Å, 2.5µm XP, 3.5µm and 5µm Columns
  • XSelect Peptide HSS T3, 100Å, 2.5µm XP, 5µm columns

**MaxPeak HPS (High Performance Surfaces) column technology.

ADDITIONAL INFORMATION

Peptides should not adsorb to the stainless steel frits or column body when using MaxPeak Premier columns.

BEH C18 is a good, general purpose, high ligand density C18 packing.

CSH C18 is a high ligand density C18 which also has positive surface charges in between the C18 ligands.  This reduces tailing of basic peptides when running in a low pH , low ionic strength mobile phase such as 0.1% formic acid.

HSS T3 is a low ligand density C18 which is designed for better retention of polar peptides.  This column can run in 100% aqueous mobile phase. 

MaxPeak Premier Columns Web Page

Peptide Reversed-Phase Columns Web Page

MaxPeak Premier Columns with MaxPeak High Performance Surfaces - Ordering Information

ACQUITY PREMIER Columns for Peptide Analysis – Infographic

Maximizing Resolution in Peptide Mapping Analysis Using UPLC with a 300-mm Column

Increasing Peak Capacity in Reversed-Phase Peptide Separations with Charged Surface Hybrid (CSH) C18 Columns

Peptide Mapping and Small Protein Separations with Charged Surface Hybrid (CSH) C18 and TFA-Free Mobile Phases

Selecting a Reversed-Phase Columns for the Peptide Mapping Analysis of a Biotherapeutic Protein

Application of Difluoroacetic Acid to Improve Optical and MS Performance in Peptide LC-UV/MS

Peptide Separations Application Notebook

Peptide Columns, Reagent, and Standards Ordering Information Guide

Bioseparations and Analyses Catalog

Not able to find a solution? Click here to request help.