Can mass overload on the column cause peak fronting? - WKB71893
Article number: 71893
ENVIRONMENT
- Column conditions
ANSWER
Yes, mass overload on a column can cause fronting.
Solutions include the following:
- Reduce the sample concentration.
- Reduce the sample injection volume.
- Use a longer or wider ID column.