Is it OK to backflush an ACQUITY UPLC HSS column? - WKB68689
Article number: 68689
ENVIRONMENT
- ACQUITY UPLC
- HSS
- column
- cleaning
- reverse
- backflush
- care and use
ANSWER
Yes, at a slow flow rate.
ADDITIONAL INFORMATION
When backflushing:
- If you are using a guard column, disconnect it.
- Disconnect the column and install in the reverse direction.
- Do not connect the column to your detector.
- Flow to waste.
- Set flow rate to 0.1 mL/min for 2.1-mm ID columns and 0.3 mL/min for 3.0-mm ID columns.
See also: ACQUITY UPLC HSS Columns Care and Use Manual.