Can I connect a nanoEase M/Z ZenFit connection directly to a Thermo Ultimate 3000 nano system that has two VICI valves? - WKB67869

Article number: 67869

ENVIRONMENT

  • nanoEase M/Z
  • ZenFit connections

ANSWER

No

ADDITIONAL INFORMATION

Because of dimensional tolerance differences, installing a nanoEase M/Z column (with ZenFit fittings) directly into non-Waters V-detail (10-32, 1/16-inch OD tubing) ports is not recommended .

Instead, connect a short piece of nanoViper fittings tubing between a VICI port and the inlet of a ZenFit union (700011519). The inlet to the union connects with nanoViper tubing. Next, connect the nanoEase M/Z trap or column (with ZenFit fittings) to the outlet of the ZenFit union.

Note: Part 700011519 may not be orderable on Waters.com, You may need to order through Customer Service.

Also see nanoEase M/Z Columns Ordering Information.