What is the recommended flow rate when converting a Styragel GPC column from one solvent to another? - WKB65833

Article number: 65833

ENVIRONMENT

  • Gel permeation chromatography (GPC)
  • Styragel GPC
  • Solvent conversion

ANSWER

Flow 7.8-mm ID Styragel columns at 0.2 mL/min and 4.6-mm ID Styragel columns at 0.05-0.07 mL/min.

ADDITIONAL INFORMATION

The slow flush minimizes pressure and viscosity changes that can damage this material.