Skip to main content
Waters

What are the expected back pressures on BioResolve SCX mAb columns? - WKB65620

Article number: 65620

ENVIRONMENT

  • BioResolve SCX mAb columns
  • Strong Cation Exchange columns

ANSWER

SALT GRADIENTS:

Pressuress obtained on a Waters Alliance HPLC using a salt gradient on BioResolve SCX mAb Columns  
Column Dimension Flow Rate (mL/min) System Pressure (psi) Pressure with column (psi) Column pressure (psi)
4.6x50mm 0.3 28 1185 1157
4.6x50mm 0.4 30 1568 1538
4.6x50mm 0.5 37 1950 1913
4.6x50mm 0.6 45 2335 2290
4.6x50mm 0.7 55 2733 2678
4.6x50mm 0.8 64 3128 3064
4.6x50mm 0.9 71 3541 3470
4.6x50mm 1.0 79 3954 3875
4.6x50mm 1.1 87 4375 4288
4.6x50mm 1.2 96 4805 4709
4.6x100mm 0.3 28 2558 2530
4.6x100mm 0.4 30 3362 3332
4.6x100mm 0.5 37 4147 4110
4.6x100mm 0.6 45 4940 4895

 

Pressures Obtained on an ACQUITY UPLC H-Class Bio System using a slat gradient (constant MES Buffer pH with Increasing Salt Gradient.  Highest pressure of the gradient is with 20mM MES 10mM Salt pH 7.0, column at 30° C)

Column Dimension

Flow Rate (mL/min)

System Pressure

Pressure with column (psi)

Column Pressure (psi)

2.1x50mm

0.1

407

1934

1527

2.1x50mm

0.2

549

3560

3011

2.1x50mm

0.3

695

5090

4395

2.1x50mm

0.4

836

6734

5898

2.1x50mm

0.5

976

8435

7459

2.1x50mm

0.6

1117

10195

9078

2.1x50mm

0.7

1255

12027

10772

2.1x50mm

0.8

1397

13927

12530

2.1x100mm

0.1

407

3674

3267

2.1x100mm

0.2

549

6819

6270

2.1x100mm

0.3

695

9850

9155

2.1x100mm

0.4

836

13113

12277

 

pH GRADIENTS:

Pressuress obtained on a Waters Alliance HPLC using a Waters' pH gradient with CX buffers and BioResolve SCX mAb Columns
Column Dimension Flow Rate (mL/min) System Pressure (psi) Pressure with column (psi) Column pressure (psi)
4.6x50mm 0.3 25 1208 1183
4.6x50mm 0.4 31 1610 1579
4.6x50mm 0.5 39 2013 1974
4.6x50mm 0.6 47 2420 2373
4.6x50mm 0.7 51 2833 2782
4.6x50mm 0.8 60 3256 3196
4.6x50mm 0.9 68 3680 3612
4.6x50mm 1.0 75 4108 4033
4.6x50mm 1.1 83 4563 4480
4.6x100mm 0.3 25 2583 2558
4.6x100mm 0.4 31 3420 3389
4.6x100mm 0.5 39 4252 4213
4.6x100mm 0.6 47 5070 5023

Pressures Obtained on an ACQUITY UPLC H-Class Bio System WITH BioResolve CX-pH 10X Concentrates  (pH Gradient).  (The highest pressure is at about 30:70 A:B)

Column Dimension

Flow Rate (mL/min)

System Pressure

Pressure with column (psi)

Column Pressure (psi)

2.1x50mm

0.1

398

2080

1682

2.1x50mm

0.2

540

3875

3335

2.1x50mm

0.3

685

5579

4894

2.1x50mm

0.4

820

7326

6506

2.1x50mm

0.5

961

9081

8120

2.1x50mm

0.6

1107

10867

9760

2.1x50mm

0.7

1255

12660

11405

2.1x50mm

0.8

1403

14620

13217

2.1x100mm

0.1

398

3668

3270

2.1x100mm

0.2

540

6842

6302

2.1x100mm

0.3

685

9779

9094

2.1x100mm

0.4

820

12515

11695

 

Not able to find a solution? Click here to request help.