Can HPLC (High Performance Liquid Chromatography) columns be run on an FPLC (Fast Protein Liquid Chromatography) system? - WKB63371
Article number: 63371
ENVIRONMENT
- HPLC
- FPLC
- Column use
ANSWER
Yes, as long as the HPLC column back pressure does not exceed the back pressure limit of the FPLC system.
ADDITIONAL INFORMATION
To estimate HPLC column back pressure, download the Columns Calculator and enter your method conditions. Note that the downloadable file is an executable file. Verify with your IT administrator that it is acceptable to download an executable file.