What is a good column choice for the USP monograph for sucralfate (assay), which calls for a 3.9-mm × 30-cm column that contains packing L8? - WKB57532

Article number: 57532

ENVIRONMENT

  • USP
  • Monograph
  • Sucralfate
  • uBondapak NH2

ADDITIONAL INFORMATION

This is the specific column the USP used to validate the assay for sucralfate. To find this information on the USP-NF website, select Resources, selecting Chromatographic Column, and then search for the monograph for sucralfate.

A possible alternate column, which can maintain the L/Dp ratio compared to the uBondapak NH2 column is:

SKU: PSS831113 - Spherisorb Amino (NH2) Column, 80Å, 5 µm, 4.6 mm X 150 mm, 1/pk

Note: Changing columns for USP monographs me result in different selectivity and/or retention times of analytes of interest.