How do I connect Waters Spherisorb 30-mm guard cartridges to a Waters Spherisorb analytical column? - WKB56162

Article number: 56162

ENVIRONMENT

  • Waters Spherisorb HPLC columns
  • Guard cartridges

ANSWER

The Waters Spherisorb 30-mm cartridges require the following end connectors in order to connect to an analytical column (which has end fittings):

PSS614100 - REMOVABLE COLUMN END FITTING (2/PK)

Once the end fittings are screwed onto the 30-mm guard cartridge, the following options enable you to connect to your analytical column:

1. Use a short length of precut tubing with 1/16" OD and install two of 186008714 - PEEK Finger-tight Fitting, 10-32, 1 ea (formerly PSL613315). 

clipboard_ee27a1f284070fc744291cf0f1018f6af.png

2. Purchase one of the following connectors:

700009535 - ASSY, TUBE, .004 ID, COL. CONN., "U"  

clipboard_e5d88f61d9a2abb177db800de139afa0b.pngFerrules are not staked on tubing upon receipt.  
The two-piece ferrule is permanently seated once the fitting is tightened into the column.
 

186006613 - .005"x1.75" UPLC Connection Tubing, 2 PK

clipboard_e16e80c2ab0490d861a61c8266d15f76c.pngFerrules are lightly staked on tubing upon receipt.
The two-piece ferrule is permanently seated once the fitting is tightened into the column.
 

700009524 - ASSY, TUBE, .005 ID, COL. CONN., INLINE

clipboard_e19936e5e10903a86d7837a3f2cb9aea7.pngFerrules are lightly staked on tubing upon receipt.
The two-piece ferrule is permanently seated once the fitting is tightened into the column.
 

ADDITIONAL INFORMATION

 