What is the difference between ACQUITY UPLC BEH C18, ACQUITY UPLC Peptide BEH C18 and ACQUITY Premier Peptide BEH C18 columns?- WKB52711
ENVIRONMENT
- ACQUITY
- ACQUITY Premier
- MaxPeak
- High Performance Surfaces
- BEH C18
- Peptide BEH C18
- Column
ANSWER
All three designations refer to the same particle technology (bridged ethylene hybrid). The Differences are as follows:
ACQUITY UPLC BEH C18: (130A)
- This type of column is packed in traditional stainless-steel hardware (frits and column body).
- The bulk packing used in this column is first QC tested with a small molecule test mix to ensure batch to batch reproducibility.
ACQUITY UPLC Peptide BEH C18 (130A and 300A)
- This type of column is packed in traditional stainless-steel hardware (frits and column body).
- The bulk packing used in this column is first QC tested with a peptide test mix (cytochrome C tryptic digest) to ensure batch to batch reproducibility for peptide separations.
ACQUITY Premier Peptide BEH C18 (130A and 300A)
- This type of column is packed in MaxPeak High Performance Surfaces (HPS) hardware.
- Very inert column frits and column body.
- Better peptide recovery out of the box.
- The bulk packing used in this column is first QC tested with a peptide test mix (cytochrome C tryptic digest) to ensure batch to batch reproducibility for peptide separations.
ADDITIONAL INFORMATION
Hybrid Particle Technology: Part 2: BEH: White Paper
ACQUITY UPLC and ACQUITY Premier BEH Columns Care and Use Manual
ACQUITY UPLC and ACQUITY Premier Peptide BEH C18, 130 Å and 300 Å Columns Care and Use Manual
MaxPeak Premier | Non-Specific Adsorption – Infographic
ACQUITY Premier Columns for Peptide Analysis – Infographic
Low Adsorption HPLC Columns Based on MaxPeak High Performance Surfaces – White Paper
For additional information on the Cytochrome C Digestion Standard (186006371), see the Cytochrome C Digestion Standard Care and Use Manual.