What is the compression fluid or gas used in Waters Radial Compression Modules? - WKB51977

Article number: 51977

ENVIRONMENT

  • Radial Compression Modules (RCMs)
  • 8 x 100 mm Cartridge Holder 
  • PrepLC 25 mm Module
  • PrepLC Assembly
  • PrepPak 1000 Module

ANSWER

Suggested fluid or gas:
8mm x 100mm Cartridge Holder: 95:5 water/methanol or 95:5 water:isopropanol
PrepLC 25 mm Module: 100% water
Prep LC Assembly 25 mm x 100 mm: 100% isopropanol
Prep LC Assembly 40 mm x 100 mm: 100% isopropanol
PrepPak 1000 Module: nitrogen gas

Also see: Where can I find the Care and Use manuals (User Guides) for Waters Radial Compression Modules (Holders)?