What is the largest inner diameter tubing size that is used on an LC system when reducing system dispersion to attach a 2.1-mm ID column? - WKB50618

Article number: 50618

ENVIRONMENT

  • HPLC system tubing

ANSWER

The largest inner diameter tubing that is used to optimize system dispersion and enhance the performance of a 2.1-mm ID column is 0.005 inch.

ADDITIONAL INFORMATION