How do I connect the VanGuard Pre-column to an ACQUITY column? - WKB48031
Article number: 48031
OBJECTIVE or GOAL
Connect VanGuard Column Protection to an ACQUITY UPLC column and minimize any void space in the connection.
ENVIRONMENT
- VanGuard Column Protection
- ACQUITY UPLC
- BEH 1.7um
PROCEDURE
- Carefully remove the black O-ring from the ferrule. (The ferrule is only lightly staked.).
- Do NOT pick up the VanGuard Pre-column by the compression screw after O-ring removal. Doing so, will cause the 2-piece ferule and the compression screw to disconnect from the tuning.
- Hold the VanGuard Pre-column with the inlet side of the VanGuard on the bench top and the ferrule end pointing up, away from the bench top.
- Bring the column inlet down over the ferrule end of the VanGuard Pre-column.
- While the inlet of the column is facing down, the tighten the compression screw until it is finger-tight.
- Always maintain a little pressure on the VanGuard Pre-column during this process to make sure the tubing goes all the way into the column inlet end nut properly.
- Flip the column/VanGuard stack so the VanGuard Cartridge is on top of the column.
- Use two wrenches to further tighten the VanGuard Pre-column to the inlet of the column.
- Tighten the compression screw of the VanGuard Pre-column about 1/4 inch turn past finger-tight. You may need to tighten slightly more than 1/4 inch turn if a leak is present after 1.4 inch turn.
- It is normal to see some threads.
- Do not over tighten.
- Use two wrenches to further tighten the VanGuard Pre-column to the inlet of the column.
- Remove the VanGuard Pre-column if you would like to check that the ferrule is seated properly.
ADDITIONAL INFORMATION
VIDEO on Waters.com How to Properly Install a VanGuard Pre-column onto an ACQUITY UPLC Column
VanGuard Column Protection Products Care and Use Manual
