Can I connect a nanoEase M/Z column with a ZenFit connection directly to the Thermo 1200 EZ nano system? - WKB32494
Article number: 32494
ENVIRONMENT
- nanoEase M/Z
- nanoEase
- ZenFit
- nanoViper
- Thermo
ANSWER
No. An additional union is required.
ADDITIONAL INFORMATION
To connect nanoEase M/Z 75um ID, 100um ID and 150um ID columns:
- Use one ZenFit union, part number SKU: 700011519 - ZenFit Union, to make connections.
- Connect a short piece of nanoViper tubing from the "nanoViper only" port on the injector valve to SKU: 700011519 - ZenFit Union.
- Then, connect nanoEase M/Z columns to the opposite side of the union.
To connect nanoEase M/Z trap columns:
- Use the two ZenFit unions, SKU: 700011519 - ZenFit Union, to make connections.
- Connect a short piece of nanoViper tubing from the "nanoViper only" port on the injector valve to SKU: 700011519 - ZenFit Union.
- Then, connect nanoEase M/Z trap column to the opposite side of the union.
- Connect a short piece of nanoViper tubing from the "nanoViper only" port on the trap valve to SKU: 700011519 - ZenFit Union.
- Then, connect nanoEase M/Z trap column to the opposite side of the union.
The inlet to the union connects with nanoViper tubing (with nanoViper fittings), and the outlet of the union connects to the nanoEase M/Z 75um, 100um and 150um ID columns (ZenFit fittings) as well as nanoEase M/Z trap columns (ZenFit fittings).
For additional information about nanoEase M/Z ZenFit columns, see nanoEase M/Z Columns Ordering Information.