Article number: 29097

OBJECTIVE or GOAL

Measure column plate count.

ENVIRONMENT

  • Column
  • plate count
  • GPC
  • APC
  • HPLC
  • ACQUITY

PROCEDURE

  1. Follow the column efficiency test procedure as stated on the test sheet.
  2. Measure the peak width at the % height specified by the plate count calculation the column test procedure used to calculate.
  3. Conduct this procedure on a column that has not been used, so it can serve to benchmark column performance over time.

ADDITIONAL INFORMATION

For additional information, see Figure 6 in the Styragel Columns Care and Use Manual

For additional information also see Figure 10 on web page Bands, Peaks and Band Spreading and Analytical Columns and Cartridges Care and Use Manual

Thi picture is an example of methods of Calculating Column Efficiency; it illustrates the use of the different peak widths of a Gaussian peak for the calculation of column efficiency (N).

Plate count peak and formula.JPG