Article number: 28254

ENVIRONMENT

  • XBridge
  • ACQUITY
  • Amide column

ANSWER

Flush the column with 50 column volumes of 60:40 acetonitrile/water with your buffer to partition water on the surface.

ADDITIONAL INFORMATION

Amide comes shipped in 100% acetonitrile and must be conditioned prior to equilibration.

For additional information, see the XBridge Columns Care and Use Manual, page 10, or the XBridge XP 2.5 µm Columns Care and Use Manual, page 11.

For ACQUITY, see the ACQUITY UPLC BEH Columns Care and Use Manual, page 11.