What do I need to do to start using my XBridge or ACQUITY Amide column? - WKB28254
Article number: 28254
ENVIRONMENT
- XBridge
- ACQUITY
- Amide
- BEH
- start up
- flush
- column
- care and use
ANSWER
Flush the column with 50 column volumes of 60:40 acetonitrile/water with your buffer to partition water on the surface.
ADDITIONAL INFORMATION
Amide comes shipped in 100% acetonitrile and must be exposed to enough water prior to equilibration, for best attraction of polar compounds to the surface.
For additional information, see the XBridge Columns Care and Use Manual, page 10, or the XBridge XP 2.5 µm Columns Care and Use Manual, page 11.