How do I translate the 14-digit serial number for Waters columns? - WKB26147
ENVIRONMENT
- HPLC columns
- UPLC columns
- serial number
- translate
- decipher
- read
- information
- ACQUITY
- BEH
- CSH
- Symmetry
- Xterra
- Resolve
- XBridge
- Sunfire
- BioResolve
ANSWER
The standard Waters Column serial numbering system that is used for columns like XBridge, Symmetry and XTerra HPLC columns as well as Acquity UPLC columns is as follows:
The serial number is a 14-digit number. (Note: The first 10 digits of the serial number are the column lot number as well.)
Part# WAT066224 - SYMMETRY C18 3.5U 4.6X75 COL, Serial #: 02233724923829
Digits 1, 2, 3, and 4 = Batch number (gel lot number). (0223 = batch 223)
Digit 5 = Manufacturing location. (3 = Wexford, Ireland; 1 = Taunton, MA)
Digit 6 = last digit of year (7 = 2017)
Digits 7, 8, and 9 = day of year (day of the year - in this case day 249 of year 2017 = 06-SEP-2017 Date of manufacture)
Digit 10 = Production run that this column came from (2 = this column was packed during production run number two on the date of manufacture)
Digits 11 and 12 = Packing station number (38 = packing station #38)
Digits 13 and 14 = Column's position in that production run (29 = the 29th column packed during the production run)
ADDITIONAL INFORMATION
Also see: Can I determine what a column description is based only on a serial number?