Can I run 100% aqueous mobile phase through an ACQUITY UPLC BEH C18, CSH C18, or HSS C18 column? - WKB258653
Article number: 258653
ENVIRONMENT
- ACQUITY UPLC
- Bridged Ethylene Hybrid (BEH)
- High Strength Silica (HSS)
- Charged Surface Hybrid (CSH)
- C18
- Mobile phase
- Dewetting
ANSWER
Yes, you can run 100% aqueous phase through a BEH C18, CSH C18, or HSS C18 column.
ADDITIONAL INFORMATION
- If you shut the system down, it is recommended that organic be introduced through the column to help rewet the pores prior to running 100% aqueous.
- A BEH C18, CSH C18, or HSS C18 column must not sit in or be stored in 100% aqueous mobile phases.
- They should be flushed with at least 20% organic every day.
- If they will not be used for four days or more, they must be stored in 100% acetonitrile.
ACQUITY UPLC and ACQUITY Premier BEH Columns Care and Use Manual
ACQUITY UPLC and ACQUITY Premier CSH Columns Care & Use Manual
ACQUITY UPLC and ACQUITY Premier HSS Columns Care and Use Manual