ANSWER

Waters does not sell MaxPeak Premier Guard Cartridges for MaxPeak Premier columns that are not of the VanGuard FIT format.

ADDITIONAL INFORMATION

VanGuard Pre-Columns may be used as guards to protect ACQUITY Premier BEH C18, 2.1mm I.D. columns. If analytes are likely to adsorb to stainless steel, several analyte injections may be needed to passivate the traditional stainless steel VanGuard Pre-column before acceptable analyte recovery is seen.